Technical Note: Noble gas extraction procedure and performance of the Cologne Helix MC Plus multi-collector noble gas mass spectrometer for cosmogenic neon isotope analysis

Benedikt Ritter, Andreas Vogt, and Tibor J. Dunai

Viewed

Total article views: 2,582 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
1,420 1,085 77 2,582 65 70
  • HTML: 1,420
  • PDF: 1,085
  • XML: 77
  • Total: 2,582
  • BibTeX: 65
  • EndNote: 70
Views and downloads (calculated since 26 Apr 2021)
Cumulative views and downloads (calculated since 26 Apr 2021)

Viewed (geographical distribution)

Total article views: 2,582 (including HTML, PDF, and XML) Thereof 2,407 with geography defined and 175 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 

Cited

Latest update: 31 May 2025
Download
Short summary
We describe the design and performance of a new noble gas mass laboratory dedicated to the development of and application to cosmogenic nuclides at the University of Cologne (Germany). At the core of the laboratory are a state-of-the-art high-mass-resolution multicollector Helix MCPlus (Thermo-Fisher) noble gas mass spectrometer and a novel custom-designed automated extraction line, including a laser-powered extraction furnace. Performance was tested with intercomparison (CREU-1) material.