Articles | Volume 3, issue 2
https://doi.org/10.5194/gchron-3-421-2021
https://doi.org/10.5194/gchron-3-421-2021
Short communication/technical note
 | 
19 Aug 2021
Short communication/technical note |  | 19 Aug 2021

Technical Note: Noble gas extraction procedure and performance of the Cologne Helix MC Plus multi-collector noble gas mass spectrometer for cosmogenic neon isotope analysis

Benedikt Ritter, Andreas Vogt, and Tibor J. Dunai

Viewed

Total article views: 2,316 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
1,271 974 71 2,316 53 56
  • HTML: 1,271
  • PDF: 974
  • XML: 71
  • Total: 2,316
  • BibTeX: 53
  • EndNote: 56
Views and downloads (calculated since 26 Apr 2021)
Cumulative views and downloads (calculated since 26 Apr 2021)

Viewed (geographical distribution)

Total article views: 2,316 (including HTML, PDF, and XML) Thereof 2,152 with geography defined and 164 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 

Cited

Latest update: 18 Nov 2024
Download
Short summary
We describe the design and performance of a new noble gas mass laboratory dedicated to the development of and application to cosmogenic nuclides at the University of Cologne (Germany). At the core of the laboratory are a state-of-the-art high-mass-resolution multicollector Helix MCPlus (Thermo-Fisher) noble gas mass spectrometer and a novel custom-designed automated extraction line, including a laser-powered extraction furnace. Performance was tested with intercomparison (CREU-1) material.