Articles | Volume 7, issue 3
https://doi.org/10.5194/gchron-7-229-2025
https://doi.org/10.5194/gchron-7-229-2025
Research article
 | 
16 Jul 2025
Research article |  | 16 Jul 2025

Environmental gamma dose rate measurements using cadmium zinc telluride (CZT) detectors

Sebastian Kreutzer, Loïc Martin, Didier Miallier, and Norbert Mercier
Abstract

The accurate and precise determination of the environmental dose rate is pivotal in every trapped-charge dating study. The environmental γ dose rate component can be determined from radionuclide concentrations using conversion factors or directly measured in situ with passive or active detectors. In-field measurements with an active detector are usually inexpensive and straightforward to achieve with adequate equipment and calibration. However, despite the rather widespread use of portable NaI or LaBr3 scintillator detectors, there is a lack of research on the performance and practicality of portable alternative detectors in dating studies, particularly in light of newer developments in the semi-conductor industry. Here, we present our experience with two small portable semi-conductor detectors housing cadmium zinc telluride (CZT) crystals. Given their small volume and low power consumption, we argue they present attractive alternatives for γ dose rate measurements in dating studies. Despite high relative detection efficiency, their small volume may pose different challenges, resulting in impractical measurements in routine studies, and therefore need investigation. In our study, we simulated the particle interaction of the CZT crystal with GEANT4 in different sediment matrices to quantify the energy threshold in the spectrum above which the count/energy count rate correlates with the environmental gamma dose rate irrespective of the origin of the γ photons. We compared these findings with experimentally derived cumulative spectra and dose rate calibration curves constructed from reference sites in France and Germany, which yielded unrealistically low threshold values, likely due to the limited variability of the investigated sites. We additionally report negligible equipment background and required minimal measurement time of only 20 min in typical environments. Cross-checking our calibration on a homogeneous loess deposit near Heidelberg confirmed the setting and assumed performance through a nearly identical γ dose rate of 1107 ± 65 µGy a−1 (CZT) to 1105 ± 11 µGy a−1 (laboratory). The outcome of our study gives credit to our threshold definition. It validates the similarity of the two investigated probes, which may make it straightforward for other laboratories to implement the technique effortlessly. Finally, the implementation of CZT detectors has the potential to streamline fieldwork and enhance the accuracy and precision of trapped-charge dating-based chronologies.

Share
1 Introduction

Assessing the effective environmental dose rate is indeed crucial for accurate and precise ages in luminescence and electron-spin-resonance (short: trapped-charge) dating studies. The dose rate plays a vital role in the age equation as it is a significant factor in determining the amount of radiation absorbed over time.

Field procedures typically involve sampling sufficient bulk material around the sampling site. The material is then analysed to quantify the natural radionuclide concentrations (such as U, Th, and K concentrations) as major contributors to the environmental radiation. If collecting sufficient material is not feasible around the sampling site, e.g. in archaeological excavations, or drilled cores, the required material can be carefully separated from the to-be-dated material combined with in situ measurements.

Strategies to ensure a good environmental dose rate estimation ideally include laboratory and field measurements. Standard analytical methods in the laboratory involve α and β counting, γ-ray spectrometry (e.g. Aitken1985; Zöller and Pernicka1989; Hutton and Prescott1992; Preusser and Kasper2001; Godfrey-Smith et al.2005; Mauz et al.2021; Kolb et al.2022), and element analytical methods such as inductively coupled plasma mass spectrometry (ICP-MS) (e.g. Preusser and Kasper2001) or a combination of these methods. Activity or element concentrations are then converted for each type of radiation (α, β, and γ radiation) using dose rate conversion factors (latest update: Cresswell et al.2018).

Dose rate (γ and, more challenging, β) components can be measured in the field at the sampling position using passive dosimeters (e.g. Hutton and Prescott1992; Kalchgruber et al.2003; Kalchgruber and Wagner2006; Richter et al.2010; Kreutzer et al.2018b) stored over a couple of weeks to months or with active detectors (usually γ-ray probes) (e.g. Murray et al.1978; Mercier and Falguères2007; Guérin and Mercier2011; Arnold et al.2012; Bu et al.2021; Martin et al.2024), enabling nearly instant dose rate estimates.

Regardless of the preferred method and type of detector, active or passive, in-field measurements are indicated when sampling suggests a heterogeneous distribution of radionuclides or complex geometries (e.g. the close succession of very different sediment layers, gravels/rocks in the profile). The field dose rates can later be compared to laboratory results based on the radionuclide concentrations. Ideally, the obtained numbers statistically agree, or the discrepancy gives further insight into the site's matrix composition. Active detectors can be paired with a portable luminescence reader (e.g. Sanderson and Murphy2010) to profile the stratigraphy and determine relative chronologies. A rule of thumb would approximate the γ dose component as about 28 % to 36 % of the total dose rate (numbers derived from the ChronoLoess database by Bosq et al.2023; alternatively, see estimates in Aitken1985). These numbers underpin the importance of the γ dose rate contribution and its significance in estimating accurate trapped-charge ages.

On the flip side, the usually short measurement durations, compared to the expected age of the sediment, have the disadvantage that long-term changes in the water content are not reflected. Conversely, passive dosimeters would register at least seasonal variations if stored over months in the field. Both (passive and active) do not register potential radioactive disequilibria. Furthermore, depending on the size of the detector probe, a rather large hole with a depth of at least ca. 30 cm is required for the measurement. Such a hole is sometimes difficult to dig, not always possible (samples from a drilled core), or not favoured given the setting (e.g. archaeological excavation). Here focussing on active detectors, additional everyday challenges involve equipment calibration and handling usually proprietary hardware such as cables or multi-channel analysers that are costly to repair or even unavailable after they have been phased out by the manufacturer. Lastly, the equipment can be bulky, especially for large detectors (up to 3 in.× 3 in. for a portable NaI probe), and, given first-hand experience, the equipment is prone to preferred inspection during air travel and requires an export port licence due to being dual-use and cannot be brought into every country.

In summary, while in-field measurements with active detectors do present certain challenges, their benefits are still considerable. They provide valuable, real-time data at a relatively low cost, significantly improving the accuracy of dating studies. As a result, their routine use seems advisable.

In the following, we will test two commercially available portable cadmium zinc telluride (CZT) detectors for in situ γ-ray measurements. The detectors are small and highly portable, and we assume that they can pose an alternative to systems using larger NaI or LaBr3 probes in trapped-charge dating applications. Next, we will begin outlining the technical specifications and advantages of the CZT systems. We will then detail the required calibration methods and explore the performance and dose-response characteristics of the detectors through simulations and measurements in different natural sites with well-known radionuclide concentrations. Finally, we will test the calibrated systems in a loess deposit near Heidelberg and discuss the results.

In this contribution, we focus exclusively on the “threshold” technique (Løvborg and Kirkegaard1974, further details below) for measuring environmental γ dose rates (D˙γ in  µGy a−1). Unlike the “window” method (three windows each for U, Th, and K), which compares the area under a specific γ peak in a sample with unknown composition to the area of a γ peak in a sample with known radionuclide composition, the threshold technique integrates the entire spectrum above a set threshold. The threshold approach provides a direct measure of D˙γ in  µGy a−1 rather than a radionuclide concentration for natural environments.

2 Material and methods

2.1 Brief background γ detectors

Measuring γ rays translates to observing the interaction of (γ) photons with matter by quantifying the production of secondary charged particles. Suitable are scintillation detectors such as NaI(Tl) or LaBr3, collecting light caused by the interaction of the γ photons with the detector material. Alternatively, semiconductor-based detectors (e.g. high-purity Ge), combined with suitable electronics, record the number of produced secondary hole pairs (e.g. Gilmore2008).

To measure γ rays outside a laboratory, for instance, in trapped-charge dating studies, portable detectors that can be operated at room temperature are preferred. This usually favours scintillation detectors using NaI(Tl) or LaBr3, with typical probes ranging from 1.5 in.× 1.5 in. to 3 in.× 3 in. over HPGe-semiconductor-based detectors that require operation at liquid nitrogen temperature due to the small band gap of the crystal. Cadmium zinc telluride (CdZnTe; short: CZT) detectors were proposed as promising alternatives with better γ-ray absorption performance and operation at room temperature. However, the production process is more challenging (e.g. Gilmore2008), and such detectors have not been an option considered in the context of trapped-charge geochronology yet.

Since the 1990s, the development of CZT semiconductor detectors has progressed considerably in their applicability as γ-ray detectors (for reviews, see Scheiber and Chambron1992; Verger et al.1997; Limousin2003; Alam et al.2021). They offer a small volume and operate at ambient temperature by collecting charges created by the interaction of ionising radiation with a high relative efficiency for photoelectric interaction (atomic numbers Cd: 48, Te: 52; density crystal ca. 5.8 g cm−3) (Limousin2003; Alam et al.2021). Although this is less important in our case, they provide an energy resolution comparable to or better than LaBr3 and considerably higher than NaI(Tl) probes (Alexiev et al.2008). Also considering the small volume available for detection, their absolute efficiency remains lower than that of larger detectors. This feature, combined with a low energy consumption, renders this detector type particularly appealing for our application.

2.2 Equipment

For our experiments, we used two systems from Kromek (https://www.kromek.com/, last access: 17 August 2024) with CZT detectors. (1) RayMon10® (henceforth: RayMon GR1) and (2) GR1+® (henceforth: GR1)1 (Fig. 1). Both systems include a similar 10 mm× 10 mm× 10 mm GR1 CZT detector connected to a 4096 energy–channel analyser. The detection ranges from 30 keV to 3 MeV with an energy resolution of around 2.5 % FWHM at 662 keV. The RayMon GR1 was delivered with a handheld touch-screen device running Microsoft Windows 10® and comes housed. The probe communicates with the handheld device via a Universal Serial Bus (USB) Type A connector. The battery lasts around 8–10 h, depending on the display brightness setting. Although much smaller in housing size (GR1: 25 mm× 25 mm× 63 mm, 60 g; compared to RayMon GR1: 42 mm× 35 mm× 180 mm, 176 g), the GR1 contains a similar CZT detector. It has a Mini-A USB port that can be attached to any standard computer given a suitable cable and operated using the software K-Spect® that can be downloaded free of charge from the manufacturer. The GR1 consumes only 250 mW and is hence operational as long as the battery of the connected computer lasts. For more information, we refer to the manufacturer's website.

https://gchron.copernicus.org/articles/7/229/2025/gchron-7-229-2025-f01

Figure 1Kromek detectors used for our measurements (shown is the probe without the handheld tablet PC for the RayMon10®). Both probes house a similar CZT detector. We wrapped the GR1® in a standard plastic bag and attached a homemade strain relief to the GR1® to enable easier operation and retraction of the detector in the field.

Download

Because the Mini-USB port of the GR1® seemed fragile, and we were not sure about the sealing of the housing against moisture, we designed a 3D-printed, rubber-sealed strain relief mount (Fig. 1) and attached it to the detector housing. The strain relief enables safe retrieval of the detector, and the simple plastic bag wrapping keeps dirt and moisture away during field operations. The adapter was designed by the Scientific Workshop Service of Heidelberg University and we share the print-ready files under CC BY-NC 4.0 licence conditions on Zenodo alongside this article.

2.3 Calibration methods

We aim to use the detectors in routine dating applications to determine D˙γ in µGy a−1. This requires three separate experiments in given order to set up each device: (1) channel–energy calibration, (2) energy threshold definition, and (3) a calibration curve modelling counts against the environmental dose rate.

The dead time was insignificant during all the experiments presented in this paper. The dead time is the difference between real time and live time, which equals the time when the detector did not register new counts. The longest dead time of both detectors was 3.6 s, with the highest relative dead time amounting to only 0.1 %.

https://gchron.copernicus.org/articles/7/229/2025/gchron-7-229-2025-f02

Figure 2Geometry of the simulated CZT detector and its placement in the sediment matrix box. (a) Image from GEANT4 visualization only showing the detector part. (b) Isometric plot showing the placement in the sediment matrix box. Both graphics not to scale. Legend: (1) sediment matrix, (2) rubber coating, (3) aluminium shielding, (4) air, (5) plastic support, (6) CZT crystal, and (7) electronic components. Technical information used for building the geometry was kindly provided by the Kromek Group plc. The dimensions of the sediment matrix box are 1.635 m× 1.635 m× 1.67 m. The production cut for the simulation was 0.5 keV. For full details on dimensions and composition, we refer to the GEANT4 code, available on Zenodo alongside this contribution.

Download

2.3.1 Channel–energy calibration

The channel–energy calibration (Sect. 3.1) assigns energy values (in keV) to the, in our case, 4096 channels. The calibration makes it easier to interpret the γ-ray spectrum, enables a comparison of spectra, and accounts for shifts in the spectrum that may occur due to, for instance, changed environmental conditions.

Both detectors used here were delivered with a test and inspection sheet documenting measurements against 241Am (γ line at 59.5 keV) and 137Cs (γ line at 662 keV). The results are nearly identical for both detectors with an offset of ca. two channels per kiloelectron volt (keV) between the GR1 and the RayMon GR1.

For the channel–energy calibrations, where only the peak position matters, we used two γ standards available in Heidelberg closely arranged around the detector for two measurements over 3600 s. One source is a homemade uranium standard (U concentration: 1.02 %) and the other an Amersham EB 165 mixed radionuclide standard with 241Am and 137Cs. The Amersham standard also contains other shorter-lived radionuclides; however, given the age of the standard (> 30 years), we do not expect to observe significant counts above the background within the chosen measurement time.

2.3.2 Energy threshold determination

The energy threshold definition (Sect. 3.5) determines the threshold in the spectrum above which D˙γ is seemingly independent of the origin of the absorbed γ photons (see Løvborg and Kirkegaard1974,  for details). In other words, the integrated spectrum above the threshold is used to derive D˙γ. Guérin and Mercier (2011) distinguished two different thresholds techniques for integrating the spectrum. The “count” and the “energy” threshold (integration technique). The count threshold adds all counts above a certain threshold (η), whereas the energy threshold integrates the deposited energy above η. Assuming that Si is the signal registered either as absolute counts per channel or count rate per channel (s−1) in the ith channel of the spectrum, Ei (in keV) is the energy associated with a certain channel. The relationship between the environmental γ dose rate and integrated value above the threshold for an energy–channel-calibrated spectrum becomes, in the case of the counting threshold technique,

(1) D ˙ γ Σ i := η N S i ,

and it reads

(2) D ˙ γ Σ i := η N S i × E i

for the energy threshold integration technique. Guérin and Mercier (2011) found η slightly lower for the latter technique, resulting in a larger proportion of the spectrum usable, which lowers the statistical uncertainty. Although related, the two threshold integration techniques must be distinguished from quantifying η, i.e. finding the energy above which D˙γ is a function of the integrated counts (Løvborg and Kirkegaard1974), regardless of the integration technique. To determine η, one can perform energy–matter interaction simulations (e.g. Guérin and Mercier2011) or measure the γ-ray spectra of “pure” emitters of known U, Th, and K concentrations (Mercier and Falguères2007; Rhodes and Schwenninger2007; Duval and Arnold2013).

For our contribution, we modelled the threshold (henceforth “ηsim”, in keV) with GEANT4 (Agostinelli et al.2003) using three different sediment matrices: (1) a pure SiO2 matrix, (2) a brick-like matrix (SiO2: 66 %, Al2O3: 18 %, Fe2O3: 6 %), and (3) a calcite-rich sediment (CaCO3: 60 %, SiO2: 40 %). We set the matrix densities to 1.8 g cm−3 and added no water (dry matrix). The simulation geometry represents the RayMon10® probe according to the available documentation provided by the manufacturer (Fig. 2). Dimensions and material of the prove were provided by the manufacturer. The probe was placed at the centre of the (1.635 × 1.635 × 1.67) m3 sediment box. In this geometry the probe is surrounded by at least 80 cm of sediment in every direction, ensuring an infinite γ-radiation matrix around it (> 99 % of the infinite matrix γ dose rate). The γ-ray emissions were simulated from each matrix using GEANT4 electromagnetic physics from the G4EmPenelopePhysics constructor (Ivanchenko et al.2011), which is based on the 2008 version of the PENELOPE Monte Carlo code for low energy particles (Baró et al.1995). This GEANT4 physics was successfully already tested for simulating γ photons from natural radionuclides in Guérin and Mercier (2011), Guérin and Mercier (2012), and Martin et al. (2015).

The γ spectra of 40K, the U series (in secular equilibrium), and the Th series (in secular equilibrium) were built from the data of the ENSDF database as of June 2014 (https://www.nndc.bnl.gov/ensarchivals/, last access: 15 September 2024) to independently simulate 1 Gy of γ dose in the matrices. In the simulation, we recorded the energy of each γ interaction with the CZT crystal, and the spectra of counts per energy channel were built for each matrix and each γ-emission spectrum. These “measured” spectra obtained by simulation were then used to create the curve of counts/deposited energy above the energy thresholds ranging from 0 to 1000 keV. We then compared the standard deviation between the 40K, U series, and Th series curves of counts above the threshold to quantify the optimal threshold for which the number of counts/energy above is proportional to the dose rate and mostly independent of the natural radionuclide composition. We did not consider dead times because we assume that, during real counting, this phenomenon has a low impact on determining the count/energy threshold.

https://gchron.copernicus.org/articles/7/229/2025/gchron-7-229-2025-f03

Figure 3Photos of all natural sites measured in this study. Panels (a)(e) are calibration sites with known radionuclide composition, and panel (f) is a loess deposit at the Weiße-Hohl near Heidelberg (Germany) used to cross-check the equipment calibrations. Details for panels (a)(d) can be found in Miallier et al. (2009) and Richter et al. (2010), and details for panels (e) and (f) are provided in the main text. The red circles mark the measurement positions (holes) for the probe. The sites BDX (d) and FLOSSI (e) are located in areas with restricted access. (d) In the basement of the Archéosciences Bordeaux laboratory at the Université Bordeaux Montaigne in Pessac (France) and (e) at the Max Planck Institute for Nuclear Physics in Heidelberg (Germany).

Download

We compare these findings with measurements at five calibration sites (Fig. 3) to derive ηexp. Four sites are located in France and three in the vicinity of Clermont-Ferrand (France) (Miallier et al.2009), and one is a homemade brick block located in the cellar of the Archéosciences Bordeaux laboratory (Richter et al.2010). Another site is a granite block (FLOSSI) located at the Max Planck Institute for Nuclear Physics (Heidelberg, Germany). The granite block was donated by the Granitwerke Leonhard Jakob KG to the Forschungsstelle Archäometrie (Günther A. Wagner) in 1991 for the purpose of having a reference site for calibrating γ-ray spectrometers. The radioelement concentration of the block was analysed with neutron activation analysis, atomic absorption, and high-resolution γ-ray spectrometry as part of the work by Rieser (1991). Although the information from this analysis was later used by others (e.g. Hossain et al.2002; Kalchgruber2002), the values were never formally published. We therefore added the CSV file with the values from Rieser (1991) to our Zenodo dataset (Kreutzer et al.2024).

In general, the investigated sites (Fig. 3a–e) have a well-known radionuclide composition from which D˙γ can be calculated to construct γ dose rate calibration curves using the two threshold integration techniques to re-evaluate η as the value where the mean square of residuals from the model reaches the lowest value. The underlying assumption of this approach is that if the threshold is set correctly, the regression line should exhibit the best fit as a non-ideal threshold should increase the residuals due to a poor fit. Insufficient model adaptation is caused by poor counting statistics (threshold too large) or in situations where the prerequisite of the technique that D˙γ is independent of the origin of the γ photons is not fulfilled (threshold too low). We will compare those values with the experimental approach of deriving the threshold, except that we do not have access to sites with pure radionuclide concentrations but will use measurements from sites Fig. 3a–e instead.

Table 1Known γ dose rates from reference sites. The dataset listed here, except for FLOSSI, is included in the R package gamma. In the original dataset, “BDX” is termed “BRIQUE”, which is the brick block in the Archéosciences Bordeaux laboratory; for clarity, we relabelled it to BDX for our analysis. The values for FLOSSI represent the central values and standard error (Galbraith and Roberts2012) of all respective analyses given in Rieser (1991). We list the results calculated with the conversion factors by Cresswell et al. (2018). Please note that data in Miallier et al. (2009) are given as total dose rate, including the cosmic dose rate contribution. Here we can neglect the cosmic dose rate contribution as we cut the spectra at 2800 keV.

Note that U and Th concentrations are in µg g−1, K is in percentage (%), and dose rates are in µGy a−1.

Download Print Version | Download XLSX

2.3.3 Dose rate calibration curves

The dose rate calibration curve (Sect. 3.6) correlates the integrated (count and energy integration technique) signal with D˙γ from the reference sites (Table 1), i.e. the response of the detector to natural γ radiation. If established, it allows us to derive an accurate estimate of D˙γ from a natural site with unknown radionuclide composition. As pointed out by Guérin and Mercier (2011), the water content will not affect the counting rate significantly, and the established value should be applicable to sites usually probed in trapped-charge dating applications.

The D˙γ values in Table 1 differ from the values reported in Miallier et al. (2009) after we recalculated them using the conversion factors compiled by Cresswell et al. (2018). Values recalculated for other conversion factors can be found in the dataset clermont_2024 contained in the R package gamma (> v1.1.0).

Please note that for establishing the calibration curves we assumed “infinite matrix” conditions that enabled us to convert the radionuclide concentrations into dose rates (e.g. Guérin et al.2012,  for a critical review of this concept).

2.4 Radionuclide determination cross-check

To validate our calibration and post-processing procedure, we recorded natural γ spectra at the Weiße-Hohl (WH2024). The site is a gully of anthropogenic origin that cut into the famous last glacial aeolian deposits near Nussloch (Germany) (e.g. Antoine et al.2001). Today, the gully is part of a hiking trail in the area and hence easily accessible. The Nussloch loess deposits are well investigated through numerous studies, and the expected D˙γ at Weiße-Hohl was about 1 Gy ka−1 (Rieser1991). What made the measurements at this particular site interesting was that loess is typically subject to past climate and chronology studies using trapped-charge dating methods and reflect an often encountered use case.

https://gchron.copernicus.org/articles/7/229/2025/gchron-7-229-2025-f04

Figure 4Energy calibration results for detector GR1. The main plot shows the raw spectrum with known γ lines marked with dashed lines. The inset displays the calibration curve applied to all subsequently shown spectra. Peak positions were found to be similar for GR1 and RayMon GR1.

Download

We recorded two spectra over 20 min with both detectors in a 32 cm deep hole and sampled about 120 g of material for subsequent radionuclide and gravimetric water content quantification. We further extracted two subsamples for radionuclide concentration analyses in Heidelberg and Bordeaux. In Heidelberg, we employed a μDose (Tudyka et al.2018; Kolb et al.2022) and a μDose+ (Tudyka et al.2024) system on the same 3 g subsample. The sample was measured more than 2 d on each system. On another 83.3 g, we performed high-resolution γ-ray spectrometry measurements in Bordeaux (Guibert and Schvoerer1991). To compare the D˙γ calculated from the radionuclide concentrations using the conversion factors by Cresswell et al. (2018). We corrected the D˙γ measured with GR1 and RayMon GR1 for the field water content (Aitken1985). As for the calibration measurements, we assumed “infinite matrix” conditions and approximated a 4π geometry.

2.5 Data and data processing

We used GEANT4 (Agostinelli et al.2003) for the threshold modelling and processed our data with R (R Core Team2024) and the packages gamma (Lebrun et al.2020; Frerebeau et al.2024) and ggplot2 (Wickham2016). The two investigated Kromek measurement systems provide export functionality for various data formats. We opted for the ASCII format .spe and added support in the function gamma::read() to the package gamma (> v1.1.0) for this study (Frerebeau et al.2024). Except additions detailed below, our workflow uses the analysis functions of the gamma package and follows the suggestions by Lebrun et al. (2020) and the tutorials that come with the gamma (> v1.1.0) R package. This also includes the steps to determine the dose rate response curve. For clarity, it should be mentioned that the gamma package internally uses the function IsoplotR::york() (Vermeesch2018) to implement a regression analysis with correlated errors of xy values that have individual uncertainties (York et al.2004).

To ensure that the figures have colour-blind-friendly colours, we used the R package khroma (Frerebeau2024), and the paper was prepared with rticles (Allaire et al.2024). A shortened version of the R code used for all the calculations, data, and calibration output is available on Zenodo (Kreutzer et al.2024) under CC BY 4.0 licence conditions in accordance with common data-sharing guidelines.

3 Results

3.1 Energy calibration

Figure 4 shows the spectrum plot of GR1 measured over 1 h. We placed our γ standards with known composition in front of the detector. The dashed lines marked the γ lines used for the channel–energy calibration. The inset draws the energy–channel-calibration curve applied subsequently to all analysed spectra. We did not apply low-level discrimination and recorded raw count values for the measurements; i.e. the count rate was calculated in the post-processing. We have chosen the measurement time to achieve a good counting statistic.

Given the nuclide composition, we expected to see typical γ lines present in the 238U decay chain on top of 241Am and 137Cs. The manufacturer also used the latter two nuclides before delivery to test the CZT detectors' performance, and hence they provide a good reference for a cross-check. We manually identified eight γ lines in our spectrum and assigned the results to the imported spectra with the gamma::energy_calibrate() function. To ease the peak identification, we started with the 241Am and 137Cs γ lines for which we have channel-to-energy references determined by the manufacturer. For instance, the manufacturer specifies finding the 241Am peak at 59.5 keV in channel number 80 (± 10 %) and the 137Cs peak at 662 keV at 880 (± 1 %). Our calibration confirmed those values with channel number 82 for 241Am 59.5 keV and channel number 877 for 137Cs 662 keV.

The same calibration was performed with the RayMon GR1 detector but with its measurement time reduced to 900 s as a cross-check. According to test data shared by the manufacturer on request (Kromek support, personal communication via e-mail, 27 September 2024), peak areas do not differ by more than 5 % to 10 % if the equipment is operated within the specified range (0–40 °C). In our case, we found that the peak positions of RayMon GR1 were virtually identical to GR1 also under different temperature conditions (ambient temperatures at Clermont Ferrand: ca. 28 °C, at FLOSSI: ca. 18 °C; data not shown). Hence, for simplicity, we applied the GR1 channel–energy calibration to all measured spectra, and all spectra shown subsequently are energy–channel-calibrated.

3.2 Background measurements

To investigate the detector's counting background, we placed the GR1 for ca. 5 h (18 060 s) in lead housing inside a low-level background environment at the PRISNA facility (Plateforme Régionale Interdisciplinaire de Spectroscopie Nucléaire en Aquitaine) near Bordeaux. The facility is a scientific platform used for low-level γ-ray spectroscopy experiments. Figure 5 illustrates that the system background is insignificant compared to typical environmental situations. The average count rate of the sum spectrum amounts to only 0.1 s−1. Given the similarity of both detectors, we applied the same background subtraction to results from both detectors.

https://gchron.copernicus.org/articles/7/229/2025/gchron-7-229-2025-f05

Figure 5Background measurements with GR1 in the lead housing for more than 5 h. The system background is negligible compared to typical measurements in the field.

Download

https://gchron.copernicus.org/articles/7/229/2025/gchron-7-229-2025-f06

Figure 6Channel–energy-calibrated spectra as recorded in the reference sites as count rate against energy. (a) Spectra measured with detector GR1 and (b) spectra measured with detector RayMon GR1. The spectra for both detectors are virtually identical in terms of peak position and count rates. The count rate for spectra C341 is significantly lower in (a) compared to (b). This is due to a software error (see main text), and therefore this spectrum was discarded for later analysis. “PRI” refers to the background spectrum recorded in the lead housing. We limited the x axis to energy range later used for the integration: 30–2800 keV.

Download

3.3 Energy-calibrated raw spectra

Figure 6 displays all energy-calibrated γ-ray spectra measured at the sites at Clermont-Ferrand, Bordeaux, Heidelberg, and PRISNA. We show count rates instead of absolute count values to account for the different measurement time. The shortest live time was 1200 s (PEP) and the longest 18 059 s (PRI). All spectra in Fig. 6a and b are scaled and colour-coded similarly for better comparison. Site PRI (the background measurement) was only measured with the GR1.

Visible in both spectra is a dominant Compton continuum rather than distinguishable photo peaks. This observation is not surprising given the short measurement time; the low abundance of the radionuclides (e.g. Miallier et al.2009,  for the Clermont-Ferrand sites); and, of course, the relatively low absolute efficiency for the small CZT crystal. It is reassuring that all comparable raw spectra appear very similar in intensity, position, and shape, except for the C341 spectra.

The spectra recorded in site C341 (a basaltic rock) appear to show only half of the counts measured with detector GR1 compared to the RayMon GR1 detector. This discrepancy is because GR1 was controlled via an external mobile computer that went unexpectedly into sleep mode. After reactivating the computer, the software seemed to have continued counting. However, post-processing revealed that it had stopped registering γ photons. In other words, the difference between the two readings (GR1 vs RayMon GR1) for C341 is a technical error, and hence, we discarded the spectrum C341 measured with GR1 for subsequent analysis. This error can be avoided easily, but we kept it in the paper to share our experience.

3.4 Minimum required measurement time

When we performed our measurements at the reference sites, we still needed more practical experience with the two detectors. Therefore, we opted for measurement times longer than the typical setting for the LaBr3 probes (ca. 10 min). Unfortunately, hour-long measurements for one sampling spot are often impracticable, considerably reducing the practicability of in-field measurements.

To assess the reasonably required measurement time for recordings, defined as a stable count rate within uncertainties in the field, we placed the RayMon GR1 detector in the brick block at Bordeaux (site: BDX) and started measurements for 60, 300, 900, and 3600 s (Fig. 7). Given the similarity of both detectors, we assume that this experiment will also be valid for the GR1. In the post-processing we integrated all spectrum counts for the experiment using the integration settings given below and normalized them to the measurement duration. The estimated mean count rate is a little bit erratic over the first 500 s before smoothing out after 20 min of measurement time. Additional measurement time increases the count rate only slightly. We therefore conclude that 20 min suffices in typical environments to determine a reproducible signal. This time corresponds to a total number of 4370 counts.

https://gchron.copernicus.org/articles/7/229/2025/gchron-7-229-2025-f07

Figure 7Sum of counts normalized to the measurement time recorded in the brick block at Archéosciences Bordeaux. After 20 min the average count rate does not change anymore within uncertainties. The plot scales depending on the settings of η (the energy threshold), and it was arbitrarily set to 200 keV for this graph.

Download

3.5 Threshold definition

In Sect. 2.3, we outlined the concept for defining the optimal energy threshold (η) above which the count rate correlates with the absorbed dose, regardless of the nature of the emitter and the matrix composition. The threshold is, in essence, a function of particle interaction with the (CZT) detector. Løvborg and Kirkegaard (1974) estimated the energy threshold for their setup (3 × 3 in NaI detector) at 500 keV; Murray et al. (1978) settled on 450 keV for their 2 in. diameter NaI(Tl) probe. Mercier and Falguères (2007) calculated a threshold of 320 keV for their 1.5 × 1.5 in. NaI(Tl) probe, a value later largely confirmed by simulations by Guérin and Mercier (2011) (their threshold value: 296 keV). Also, Duval and Arnold (2013) reported comparable values for NaI and LaBr3 detectors of the same sizes (LaBr3)(Ce): 358 keV, NaI(Tl): 322 keV). Our unpublished observations, employing the threshold method, suggest that the threshold shifts towards higher energies for larger detectors of the same material. This phenomenon is likely attributed to the increased proportion of photons registered from 40K, relatively to photons from the U and Th series. To compensate for this larger contribution of 40K photons in the distribution, the threshold shifts towards higher energies to ensure that the total count rate is proportional to the absorbed dose. Given the small volume of our CZT detector, we would position the threshold in the low-energy portion of the spectrum, not exceeding the values reported in the literature.

https://gchron.copernicus.org/articles/7/229/2025/gchron-7-229-2025-f08

Figure 8(a, b) GEANT4 simulation results. (c, d) Cumulative γ-ray spectra for all natural calibration sites normalized to the respective environmental γ dose rate. Data only shown for RayMon GR1. (e, f) A variant of the experimental data as mean square of weighted deviates (MSWD) of the dose rate model fitting against the chosen minimum energy threshold. Values at 1 indicate the best fit. The calculated energy threshold (η) is indicated in each of the plots as a dotted line. Please keep in mind that while the three sets of graphs aim at showing the energy threshold using different methods, they represent neither the same data nor the same fitting method. For more details, see main text.

Download

3.5.1 GEANT4 simulations

Figure 8a and b exhibit the simulations results for the three different matrices. We show the relative standard deviation between the number of counts above the energy threshold recorded during the simulations of 1 Gy generated with spectra of the radionuclides of the U series, Th series, and the 40K. This standard deviation is minimized when the number of counts/energy above the threshold is less dependent on the radionuclide of a chain of origin of the γ photons, i.e. when the number of counts above the threshold is proportional to the dose absorbed by the detector, despite the origin of natural γ rays. The minimum standard deviation, obtained between 192.5 keV and 242.5 keV for the count integration technique and 97.5 and 222.5 keV for the energy integration technique, corresponds to the curves in Fig. 8a and b falling below 10 % of the relative standard deviation (horizontal dashed line in Fig. 8a and b). This represents the optimal energy range for setting the energy threshold for the detector ηsim according to our simulations for the two integration techniques, respectively.

3.5.2 Field measurements

Classical method

The “classical” method to determine ηexp experimentally is measurements in environments with different and ideally pure radionuclide compositions. Here we tried to determine the energy threshold using the calibration sites at hand for the counts and the energy integration technique (Fig. 8c and d). The threshold is defined as the smallest relative standard deviation of all spectra normalized to the respective environmental γ dose rate of the sites. In Fig. 8c and d we only show the results of the detector RayMon GR1.

For the count and energy integration technique, we obtained ηexp at 99 and at 59 keV, respectively. Both values are considerably smaller than the results from our simulation. We will show later that the simulated η is likely more accurate than the experimentally derived one. We attribute the difference to the similarity of the measured sites and to the fact that, although D˙γ varies for all sites, the U/Th ratio is rather similar, likely leading to an unrealistically low value of η.

Calibration curve fitting

As an alternative to the simulation and experimental quantification of ηexp, we experimented with a different approach. We calculated the γ dose rate response curve for different energy windows using gamma::dose_fit(). Amongst other values, the function returns the mean square of weighted deviates (MSWD), which we can use to approximate the quality of fit of our regression model. Values lower or higher than 1 indicate a poor model adaptation. We defined the moving lower energy limit as Ei (i:=30,,1000), and Emax was set to 2800 keV to avoid counts from cosmic rays. Figures 8e and f show the outcome of this calculation for both detectors (blue: GR1, red: RayMon GR1). Although the curves of the mean residuals differ, the divergence of the determined thresholds is small, and we believe that this deviation is caused by the discarded data point C341 for GR1, which is the lower point in the calibration curve.

For the count calculation technique (not shown in Fig. 8e), the minimum in the search window between 30 and 350 keV was found at 91 keV. For the energy counting calculation technique, we located the value at 71 keV. Also, these values are smaller than their simulated equivalents, and the approach cannot compensate for the lack of differences between the measured sites. We therefore decided to continue with the simulated energy threshold values (i.e. η:=ηsim).

https://gchron.copernicus.org/articles/7/229/2025/gchron-7-229-2025-f09

Figure 9Dose rate calibration curves for detector GR1 (a) and RayMon GR1 (b). Uncertainties are given as standard errors (for details, see York et al.2004). Shown is the known γ dose rate from the reference sites against the integrated energy signal between the threshold η (in keV) and 2800 keV. Note that the graphs give the impression that the uncertainties for the sites measured in France increase proportional to the size of the known γ dose rates. This is a coincidence for our data subset and not a real effect (cf. Miallier et al.2009).

Download

3.6 Dose rate calibration

With the threshold η derived from ηsim, we can obtain our dose rate model again with the function gamma::dose_fit() but this time for a fixed count/energy threshold at 99 keV and 59 keV, respectively. The results are shown in Fig. 9 for the detector GR1 (Fig. 9a) and RayMon GR1 (Fig. 9b). Visual inspection confirms a good fit of the model to the data. However, the calibration curves differ slightly between the two detectors, which is likely due to the lower number of available data points. The fitting parameters of both regression lines overlap with uncertainties (standard error as sum if weighted deviation from the fit; see York et al.2004). Acknowledging minimal variations between the CZT crystals and differences in housing and electronic, a perfect match is, however, not expected.

The gamma package automatically fits the data for the energy threshold calculation technique and the counting threshold calculation technique for a given η. In Fig. 9 we have shown only the latter. However, both values are accessible and saved in the file CAL_heiLUM_V0.rda we made accessible at Zenodo (Kreutzer et al.2024). As a reminder, we converted the radionuclide concentrations from the reference sites to dose rates using conversion factors compiled by Cresswell et al. (2018). These values influence the slope and intercept of the calibration curves. Because it may be desirable to apply additional calibrations based on other available conversion factors we repeated the calibration using conversion factors from Adamiec and Aitken (1998), Guérin et al. (2011), and Liritzis et al. (2013) (see data on Zenodo: Kreutzer et al.2024).

https://gchron.copernicus.org/articles/7/229/2025/gchron-7-229-2025-f10

Figure 10Gamma spectra recorded at Weiße-Hohl (Germany) with the two detectors. The inset shows that the absolute count rate difference between the two detectors is randomly distributed.

Download

3.7 Cross-check against natural site

The measurements at the Weiße-Hohl confirm once more that the two detectors exhibit very similar characteristics in terms of count rate efficiency (Fig. 10). Differences seem stochastic without visible systematic diversion over the measurement duration of 20 min. To estimate the uncertainties, we implemented a new routine in the gamma R package (argument: dose_predict(…,  use_MC = TRUE) that uses a Monte Carlo simulation approach, re-sampling from distributions for slope, intercept, and the signal to predict the dose rate on the regression line. We found that with this method, the uncertainty increases over the analytical approach; however, it should reflect the true uncertainty more realistically.

Table 2Dry γ dose rate results for the sample WH2024 obtained with different methods. Uncertainties on the final dose rate are quoted in 2σ approximating the 95 % confidence interval. The dose rates for the CZT estimates include an systematic error contribution of 1 % from the energy calibration and was corrected for the in situ water content. Beyond digits listed here, we calculated with the full precision as returned by the measurement devices. The CZT measurements are in situ measurements, the μDose devices sampled 3 g of material each, and the γ-ray spectrometer measurements used 88.3 g. REF_Central is the central value and its uncertainty from the laboratory-derived D˙γ values.

Note that U and Th concentrations are in µg g−1, K is in percentage (%), and dose rates are in µGy a−1. PRISNA is a γ-ray spectrometer in the Archéosciences Bordeaux laboratory. NA: not available.

Download Print Version | Download XLSX

The water content from the sample site (sample code: WH2024) was estimated at 2.1 % in the laboratory, and this value was used to correct D˙γ. Table 2 summarizes the derived dose rate results for the two threshold integration techniques. D˙γ-final is the arithmetic average of the values of these two techniques. The results for GR1 and RayMon GR1 agree within 2σ uncertainties. This observation is likely caused by the calibration of GR1 sitting on fewer data points. The comparison of D˙γ against values derived from the radionuclide concentrations on WH2024 demonstrates good agreement with field measurement uncertainties. If we compare the CZT results (GR1 and RayMon GR1) with the laboratory-derived D˙γ, both summarized as central values (e.g. Galbraith and Roberts2012), we obtained 1107 ± 65 µGy a−1 (CZT) and 1105 ± 11 µGy a−1 (laboratory).

The results indicate a good homogeneity of the site reflected in the agreement between the field and the sampling dose rate. To get a better feeling for the sensitivity of D˙γ as a function of η for our detectors, we can calculate D˙γ of WH2024 for different values η. For this experiment on the energy-calibrated spectra, we have to repeat the dose rate calibration curve fitting and then predict the new dose rate given the newly derived calibration for ηi (in keV) with i:=30,40,50,,530. The upper integration energy was set to 2800 keV analogue to the previous calculations.

https://gchron.copernicus.org/articles/7/229/2025/gchron-7-229-2025-f11

Figure 11Estimations for D˙γ for different values of η for the two investigated detectors. The solid line shows the D˙γ of the measured site Weiße-Hohl derived from laboratory radionuclide estimations and the dashed lines its uncertainties. All uncertainties are shown as 2σ values. For further details, see main text.

Download

We assume that the laboratory-derived D˙γ values are the benchmark value we want to reproduce. Figure 11 offers insight into the evolution of D˙γ for GR1 (blue) and RayMon GR1 (red). The solid line is the central value reference for WH2024 from laboratory measurements, and the dashed lines show the 2σ uncertainties (calculation according to the Central Dose Model; Galbraith and Roberts2012). Setting aside the calibration-caused discrepancy between the two detectors, both detectors perform similarly between a threshold of ca. 30 and 200 keV for the count integration technique and 30 to 150 keV for the energy integration technique and overlap before falling systematically below the laboratory-derived reference value. The uncertainties of all analyses overlap until a threshold of ca. 300 keV, and it appears that for our setting and tested environment, the output is relatively insensitive to the exact value picked for η (given the uncertainties). The picked values, however, seem to nearly ideally reproduce the benchmark value, while lower values for η as indicated in our experiments would likely overestimate the true D˙γ.

4 Discussion

Our study reported the performance and dose rate calibration procedure of two portable semiconductor-based portable γ-ray spectrometers. Both devices host a similar CZT detector that can be operated at ambient temperature, i.e. in situations typical for environmental dose rate measurements as part of trapped-charge dating studies. Unlike the literature reporting on γ-ray measurements in the field that used NaI(Tl) or LaBr3 probes with inch-size diameters, our detectors are considerably smaller (crystal volume 1 cm3), and the systems have a low power consumption, boosting their appeal for trapped-charge dating studies despite that no previous experience was available addressing our field of application. This seems surprising, given the body of available literature about CZT detectors. However, usually, those studies aim at nuclear radiation monitoring (e.g. Alam et al.2021) or identifying artificial radionuclides in environmental studies (e.g. Rahman et al.2013) for which such detectors are primarily designed.

On the plus side, this feature of the detectors simplified the energy–channel calibration with artificial radionuclides because of the detectors' sensitivity to those nuclides. Our energy calibration exhibited peak positions in excellent agreement for both detectors, and we concluded that we could apply one single energy calibration. This approach was valid for us, but other detectors likely require separate channel–energy calibration. Although we did not observe a shift in the channel–energy calibration with temperature during all experiments, we highly recommend an energy–channel calibration as part of the post-processing because all subsequent analyses depend on it.

Dating studies require an accurate reading of a sediment matrix's natural γ-radiation field of unknown radionuclide composition in a 4π geometry at the sampling position. Our study proved that both detectors can achieve this in a reasonable time of 20 min. This value likely works for many environments typically encountered in trapped-charge dating applications. Still, it might be too short for accurate dose rate estimations in settings with a low number of natural radionuclides or if higher precision is desired. Hence, in case of doubt, measurement times should be adjusted. We recommend a minimum measurement duration of 60 min to obtain the dose rate calibration curve with a good counting statistics.

A crucial part of our contribution was the determination of the energy threshold η above which the count/energy rate is proportional to the dose rate for natural radioactive elements. Given the highly comparable performance characteristics of both detectors, our results can be easily used by others with the same type of detector without repeating all experiments. We tested three different methods (simulation, classical measurements, dose rate response curve fitting) to determine this threshold and opted for the results from the simulation since the measured natural sites were likely not diverse enough. Because we had access to general schematics provided by the manufacturer, we also believe that the GEANT4 simulation should be fairly accurate. The cross-check to the natural loess site indicates that the assumption made to simplify the simulation had no significant impact on the results. The threshold found here is considerably lower than results obtained in studies with NaI(Tl) or LaBr3 probes that place η at ca. 300 keV or higher (Mercier and Falguères2007; Guérin and Mercier2011; Duval and Arnold2013). This balances to some extent the lower absolute efficiency of the tiny CZT detectors because it allows a larger portion of the recorded spectrum to be exploited.

Nonetheless, it should be kept in mind that all three approaches, simulation, classical experiments, and dose rate curve fitting, have different meanings. Under the assumption of correct input parameters, the simulation investigates the interaction of the γ photons with matter for different scenarios and can hence truly determine a range above which the threshold assumption is valid. In other words, the simulation results have merit and provide a solid basis for setting of the thresholds. On the contrary, the experiment findings depend on the matrix composition of the host rock, which, in our case, is very similar if translated into relative γ dose rate contributions from the different radionuclides. We did not observe a matching pattern for the threshold from the measurements and the simulation, and without simulation, a meaningful determination of η still requires measurements of emitters with pure radionuclide composition, such as the Oxford blocks (Rhodes and Schwenninger2007).

The threshold quantified in our study is likely not much different for detectors of similar size and with a comparable CZT detector. Therefore we argue that the threshold settings can be adapted if a simulation or a measurement is not possible. This suggestion is further supported by our tests of the Weiße-Hohl measurement with shifting thresholds, and a minor difference in the detection characteristics will not bias the outcome for the γ dose rate. Future work should investigate the calibration curve at very low dose rates and in very different environmental settings, since this was not tested in our study.

The D˙γ results of the Weiße-Hohl reflect mainly statistical variations of the different analytical methods. Striking but not puzzling is the relatively large coefficient of variation (cν) of portable CZT detector results (GR1: 11.9 %; RayMon GR1: 6.8 %) compared to the laboratory measurements (ca 1.7 % or lower). Duval and Arnold (2013) reported cνs around 5 % comparable to our laboratory measurements, and results of LaBr3 measurements calibrated at the Clermont-Ferrand sites with more data points typically yielded cνs of 5 % or better (e.g. Kreutzer et al.2018a, their Table S9) (typical values: live time: ca. 600 s; integrated counts: ca. 24 000 counts; count rate: 45 s−1)

Furthermore, the larger uncertainties of GR1 compared to RayMon GR1 seem to diminish the overall good performance. For GR1, the weaker performance (larger uncertainties) results from only having three calibration points available, which would disappear with an additional point. More generally, we argue that this precision can be significantly improved with more points for the dose rate calibration curve. Those points can be added at any time later, for instance, by measuring more sites around Clermont-Ferrand (Miallier et al.2009). In such case, however, a check on the energy calibration is mandatory before and after going to the field to monitor potential shifts of the energy spectrum that might be caused by temperature or other technical reasons. Although our results did not encounter such shift, all experiments were carried out in a very short time window.

Finally, what we did not expect of these CZT detectors but should be mentioned for completeness is that our findings show that these detectors are unsuited for applying the “window” method in environments and for measurement durations typical for our field of application. This would require hour-long measurements to achieve acceptable error margins. For the determination of radionuclide composition, laboratory-based analytical techniques are unmatched in their effectiveness and precision, and they also allow α and β dose rate components to be derived.

5 Conclusions

The primary aim of our study was to test and evaluate the performance of two portable CZT detectors for deployment as active in situ detectors in trapped-charge dating applications. To that end, we measured spectra on natural reference sites with known radionuclide composition in France to derive a dose rate calibration curves for our two detectors. Background measurements in a low-radiation setting exhibited negligible count rates that can be ignored.

To determine the optimal energy threshold above which the matrix composition of the measured site does not bias the integrated signal to γ dose rate relation, we performed energy–matter interaction simulations using GEANT4. The simulation indicated a suitable energy threshold between 192 and 242.5 keV for the count integration technique and 98 and 222 keV for the energy integration technique. We compared those with thresholds derived from cumulative spectra and signal dose rate regression lines for the two different integration techniques, and we found a value of 91 keV for the counting threshold integration and 71 keV for energy counting integration technique. However, given the results from the reference loess site with know radionuclide composition, we discarded the experimentally derived energy thresholds as they are likely too low because of the high similarity of the investigated natural sites. To record a γ dose rate in typical natural sediment environments, we recommend a measurement time of at least 20 min (this approximates to a total of 4500 counts or better).

A check of our results through measurements at the homogeneous loess deposit near Heidelberg, for which we derived the radionuclide composition in the laboratory, confirmed an excellent match of field and laboratory methods, however with considerably larger (but perhaps more realistic) uncertainties for results from the CZT detectors. Finally, we argue that refined calibrations can further reduce those uncertainties on more sites. Future work may want to extend our calibration curves and explore the performance of the detector in more extreme (low and high) natural radiation fields.

Code and data availability

Raw and partly processed data and R code used in this study are available on Zenodo (https://doi.org/10.5281/zenodo.13731839, Kreutzer et al.2024).

Author contributions

SK: writing (original draft), validation, methodology, formal analysis, conceptualization, funding acquisition, data curation, software. LM: writing (review and editing), methodology, formal analysis. DM: writing (review and editing), resources. NM: writing (review and editing), methodology, formal analysis, conceptualization.

Competing interests

The contact author has declared that none of the authors has any competing interests.

Disclaimer

Publisher's note: Copernicus Publications remains neutral with regard to jurisdictional claims made in the text, published maps, institutional affiliations, or any other geographical representation in this paper. While Copernicus Publications makes every effort to include appropriate place names, the final responsibility lies with the authors.

Acknowledgements

We thank Nicolas Frerebeau for his very responsive support as maintainer of the R package gamma. We are grateful to Maryam Heydari for performing the measurements at the Weiße-Hohl. We further thank Michael Faske from the Scientific Workshop Service of Heidelberg University for designing and printing the strain relief adapter for the GR1 detector and agreeing to share it under CC BY-NC licence conditions. Dennis Gross and Jochen Schreiner enabled the access to FLOSSI and organized the removal of the tree in front of the block. Martin Autzen and an anonymous reviewer provided us with helpful comments and suggestions. For the publication fee we acknowledge financial support by Heidelberg University.

Financial support

This research has been supported by the Deutsche Forschungsgemeinschaft (grant no. 505822867).

Review statement

This paper was edited by Julie Durcan and reviewed by Martin Autzen and one anonymous referee.

References

Adamiec, G. and Aitken, M. J.: Dose-rate conversion factors: update, Ancient TL, 16, 37–50, https://doi.org/10.26034/la.atl.1998.292, 1998. a

Agostinelli, S., Allison, J., Amako, K., Apostolakis, J., Araujo, H., Arce, P., Asai, M., Axen, D., Banerjee, S., Barrand, G., Behner, F., Bellagamba, L., Boudreau, J., Broglia, L., Brunengo, A., Burkhardt, H., Chauvie, S., Chuma, J., Chytracek, R., Cooperman, G., Cosmo, G., Degtyarenko, P., Dell'Acqua, A., Depaola, G., Dietrich, D., Enami, R., Feliciello, A., Ferguson, C., Fesefeldt, H., Folger, G., Foppiano, F., Forti, A., Garelli, S., Giani, S., Giannitrapani, R., Gibin, D., Cadenas, J. J. G., González, I., Abril, G. G., Greeniaus, G., Greiner, W., Grichine, V., Grossheim, A., Guatelli, S., Gumplinger, P., Hamatsu, R., Hashimoto, K., Hasui, H., Heikkinen, A., Howard, A., Ivanchenko, V., Johnson, A., Jones, F. W., Kallenbach, J., Kanaya, N., Kawabata, M., Kawabata, Y., Kawaguti, M., Kelner, S., Kent, P., Kimura, A., Kodama, T., Kokoulin, R., Kossov, M., Kurashige, H., Lamanna, E., Lampén, T., Lara, V., Lefebure, V., Lei, F., Liendl, M., Lockman, W., Longo, F., Magni, S., Maire, M., Medernach, E., Minamimoto, K., de Freitas, P. M., Morita, Y., Murakami, K., Nagamatu, M., Nartallo, R., Nieminen, P., Nishimura, T., Ohtsubo, K., Okamura, M., O'Neale, S., Oohata, Y., Paech, K., Perl, J., Pfeiffer, A., Pia, M. G., Ranjard, F., Rybin, A., Sadilov, S., Salvo, E. D., Santin, G., Sasaki, T., Savvas, N., Sawada, Y., Scherer, S., Sei, S., Sirotenko, V., Smith, D., Starkov, N., Stoecker, H., Sulkimo, J., Takahata, M., Tanaka, S., Tcherniaev, E., Tehrani, E. S., Tropeano, M., Truscott, P., Uno, H., Urban, L., Urban, P., Verderi, M., Walkden, A., Wander, W., Weber, H., Wellisch, J. P., Wenaus, T., Williams, D. C., Wright, D., Yamada, T., Yoshida, H., and Zschiesche, D.: GEANT4—a simulation toolkit, Nucl. Instrum. Meth. A, 506, 250–303, https://doi.org/10.1016/s0168-9002(03)01368-8, 2003. a, b

Aitken, M. J.: Thermoluminescence dating, Studies in archaeological science, Academic Press, London, ISBN 978-0-12-046381-7, 1985. a, b, c

Alam, M. D., Nasim, S. S., and Hasan, S.: Recent progress in CdZnTe based room temperature detectors for nuclear radiation monitoring, Prog. Nucl. Energ., 140, 103918, https://doi.org/10.1016/j.pnucene.2021.103918, 2021. a, b, c

Alexiev, D., Mo, L., Prokopovich, D. A., Smith, M. L., and Matuchova, M.: Comparison of LaBr3:Ce and LaCl3:Ce With NaI(Tl) and Cadmium Zinc Telluride (CZT) Detectors, IEEE T. Nucl. Sci., 55, 1174–1177, https://doi.org/10.1109/TNS.2008.922837, 2008. a

Allaire, J., Xie, Y., Dervieux, C., R Foundation, Wickham, H., Journal of Statistical Software, Vaidyanathan, R., Association for Computing Machinery, Boettiger, C., Elsevier, Broman, K., Mueller, K., Quast, B., Pruim, R., Marwick, B., Wickham, C., Keyes, O., Yu, M., Emaasit, D., Onkelinx, T., Gasparini, A., Desautels, M.-A., Leutnant, D., MDPI, Taylor and Francis, Öğreden, O., Hance, D., Nüst, D., Uvesten, P., Campitelli, E., Muschelli, J., Hayes, A., Kamvar, Z. N., Ross, N., Cannoodt, R., Luguern, D., Kaplan, D. M., Kreutzer, S., Wang, S., Hesselberth, J., and Hyndman, R.: rticles: Article Formats for R Markdown, CRAN (Comprehensive R Archive Network), https://doi.org/10.32614/CRAN.package.rticles, R package version 0.27.3, , 2024. a

Antoine, P., Rousseau, D.-D., Zöller, L., Lang, A., Munaut, A.-V., Hatté, C., and Fontugne, M.: High-resolution record of the last Interglacial-glacial cycle in the Nussloch loess-palaeosol sequences, Upper Rhine Area, Germany, Quatern. Int., 76/77, 211–229, https://doi.org/10.1016/S1040-6182(00)00104-X, 2001. a

Arnold, L. J., Duval, M., Falguères, C., Bahain, J. J., and Demuro, M.: Portable gamma spectrometry with cerium-doped lanthanum bromide scintillators: Suitability assessments for luminescence and electron spin resonance dating applications, Radiat. Meas., 47, 6–18, https://doi.org/10.1016/j.radmeas.2011.09.001, 2012. a

Baró, J., Sempau, J., Fernández-Varea, J., and Salvat, F.: PENELOPE: An algorithm for Monte Carlo simulation of the penetration and energy loss of electrons and positrons in matter, Nucl. Instrum. Meth. B, 100, 31–46, https://doi.org/10.1016/0168-583X(95)00349-5, 1995. a

Bosq, M., Kreutzer, S., Bertran, P., Lanos, P., Dufresne, P., and Schmidt, C.: Last Glacial loess in Europe: luminescence database and chronology of deposition, Earth Syst. Sci. Data, 15, 4689–4711, https://doi.org/10.5194/essd-15-4689-2023, 2023. a

Bu, M., Murray, A. S., Kook, M., Buylaert, J.-P., and Thomsen, K. J.: The Application of Full Spectrum Analysis to NaI(Tl) Gamma Spectrometry for the Determination of Burial Dose Rates, Geochronometria, 48, 161–170, https://doi.org/10.2478/geochr-2020-0009, 2021. a

Cresswell, A. J., Carter, J., and Sanderson, D. C. W.: Dose rate conversion parameters: Assessment of nuclear data, Radiat. Meas., 120, 195–201, https://doi.org/10.1016/j.radmeas.2018.02.007, 2018. a, b, c, d, e

Duval, M. and Arnold, L.: Field gamma dose-rate assessment in natural sedimentary contexts using LaBr3(Ce) and NaI(Tl) probes: A comparison between the “threshold” and “windows” techniques, Appl. Radiat. Isotopes, 74, 36–45, https://doi.org/10.1016/j.apradiso.2012.12.006, 2013. a, b, c, d

Frerebeau, N.: khroma: Colour Schemes for Scientific Data Visualization, Université Bordeaux Montaigne, Pessac, France, https://doi.org/10.5281/zenodo.1472077, R package version 1.14.0, 2024. a

Frerebeau, N., Lebrun, B., Paradol, G., and Kreutzer, S.: gamma: Dose Rate Estimation from in-Situ Gamma-Ray Spectrometry, Université Bordeaux Montaigne, Pessac, France, https://doi.org/10.32614/CRAN.package.gamma, R package version 1.1.0, 2024. a, b

Galbraith, R. F. and Roberts, R. G.: Statistical aspects of equivalent dose and error calculation and display in OSL dating: An overview and some recommendations, Quat. Geochronol., 11, 1–27, https://doi.org/10.1016/j.quageo.2012.04.020, 2012. a, b, c

Gilmore, G. R.: Practical gamma-ray spectrometry, 2nd edn., John Wiley & Sons, Ltd, ISBN 9780470861981, https://doi.org/10.1002/9780470861981, 2008. a, b

Godfrey-Smith, D. I., Scallion, P., and Clarke, M. L.: BETA DOSIMETRY OF POTASSIUM FELDSPARS IN SEDIMENT EXTRACTS USING IMAGING MICROPROBE ANALYSIS AND BETA COUNTING, Geochronometria, 24, 7–12, http://www.geochronometria.pl/pdf/geo_24/Geo24_2.pdf (last access: 19 May 2025), 2005. a

Guérin, G. and Mercier, N.: Determining gamma dose rates by field gamma spectroscopy in sedimentary media: Results of Monte Carlo simulations, Radiat. Meas., 46, 190–195, https://doi.org/10.1016/j.radmeas.2010.10.003, 2011. a, b, c, d, e, f, g, h

Guérin, G., Mercier, N., and Adamiec, G.: Dose-rate conversion factors: update, Ancient TL, 29, 5–8, https://doi.org/10.26034/la.atl.2011.443, 2011. a

Guérin, G., Mercier, N., Nathan, R., Adamiec, G., and Lefrais, Y.: On the use of the infinite matrix assumption and associated concepts: A critical review, Radiat. Meas., 47, 778–785, https://doi.org/10.1016/j.radmeas.2012.04.004, 2012. a

Guibert, P. and Schvoerer, M.: TL dating: Low background gamma spectrometry as a tool for the determination of the annual dose, International Journal of Radiation Applications and Instrumentation. Part D. Nuclear Tracks and Radiation Measurements, 18, 231–238, https://doi.org/10.1016/1359-0189(91)90117-z, 1991. a

Guérin, G. and Mercier, N.: Preliminary insight into dose deposition processes in sedimentary media on a scale of single grains: Monte Carlo modelling of the effect of water on the gamma dose rate, Radiat. Meas., 47, 541–547, https://doi.org/10.1016/j.radmeas.2012.05.004, 2012. a

Hossain, S. M., Corte, F. D., Vandenberghe, D., and den haute, P. V.: A comparison of methods for the annual radiation dose determination in the luminescence dating of loess sediment, Nucl. Instrum. Meth. A, 490, 598–613, https://doi.org/10.1016/s0168-9002(02)01078-1, 2002. a

Hutton, J. T. and Prescott, J. R.: Field and laboratory measurements of low-level thorium, uranium and potassium, International Journal of Radiation Applications and Instrumentation. Part D. Nuclear Tracks and Radiation Measurements, 20, 367–370, https://doi.org/10.1016/1359-0189(92)90066-5, 1992. a, b

Ivanchenko, V., Apostolakis, J., Bagulya, A., Abdelouahed, H. B., Black, R., Bogdanov, A., Burkhard, H., Chauvie, S., Cirrone, P., Cuttone, G., Depaola, G., Rosa, F. D., Elles, S., Francis, Z., Grichine, V., Gumplinger, P., Gueye, P., Incerti, S., Ivanchenko, A., Jacquemier, J., Lechner, A., Longo, F., Kadri, O., Karakatsanis, N., Karamitros, M., Kokoulin, R., Kurashige, H., Maire, M., Mantero, A., Mascialino, B., Moscicki, J., Pandola, L., Perl, J., Petrovic, I., Ristic-Fira, A., Romano, F., Russo, G., Santin, G., Schaelicke, A., Toshito, T., Tran, H., Urban, L., Yamashita, T., and Zacharatou, C.: Recent Improvements in Geant4 Electromagnetic Physics Models and Interfaces, Progress in Nuclear Science and Technology, 2, 898–903, https://doi.org/10.15669/pnst.2.898, 2011. a

Kalchgruber, R.: α-Al2O3:C als dosimeter zur bestimmung der dosisleistung bei der lumineszenzdatierung, PhD thesis, Heidelberg University, Heidelberg, https://doi.org/10.11588/heidok.00001965, 2002. a

Kalchgruber, R. and Wagner, G. A.: Separate assessment of natural beta and gamma dose-rates with TL from single-crystal chips, Radiat. Meas., 41, 154–162, https://doi.org/10.1016/j.radmeas.2005.04.002, 2006. a

Kalchgruber, R., Fuchs, M., Murray, A. S., and Wagner, G. A.: Evaluating dose-rate distributions in natural sediments using α-Al2O3:C grains, Radiat. Meas., 37, 293–297, https://doi.org/10.1016/S1350-4487(03)00012-X, 2003. a

Kolb, T., Tudyka, K., Kadereit, A., Lomax, J., Porȩba, G., Zander, A., Zipf, L., and Fuchs, M.: The μDose system: determination of environmental dose rates by combined alpha and beta counting – performance tests and practical experiences, Geochronology, 4, 1–31, https://doi.org/10.5194/gchron-4-1-2022, 2022. a, b

Kreutzer, S., Duval, M., Bartz, M., Bertran, P., Bosq, M., Eynaud, F., Verdin, F., and Mercier, N.: Deciphering long-term coastal dynamics using IR-RF and ESR dating: A case study from Médoc, south-West France, Quat. Geochronol., 48, 108–120, https://doi.org/10.1016/j.quageo.2018.09.005, 2018a. a

Kreutzer, S., Martin, L., Guérin, G., Tribolo, C., Selva, P., and Mercier, N.: Environmental dose rate determination using a passive dosimeter: Techniques and workflow for alpha-Al2O3:C chips, Geochronometria, 45, 56–67, https://doi.org/10.1515/geochr-2015-0086, 2018b. a

Kreutzer, S., Martin, L., Miallier, D., Mercier, N., and Faske, M.: Dataset: Environmental Gamma Dose Rate Measurements using CZT Detectors, Zenodo [data set] and [code], https://doi.org/10.5281/zenodo.13731839, 2024. a, b, c, d, e

Lebrun, B., Frerebeau, N., Paradol, G., Guérin, G., Mercier, N., Tribolo, C., Lahaye, C., and Magali, R.: gamma: An R package for dose rate estimation from in-situ gamma-ray spectrometry measurements, Ancient TL, 38, 11–15, https://doi.org/10.26034/la.atl.2020.544, 2020. a, b

Limousin, O.: New trends in CdTe and CdZnTe detectors for X- and gamma-ray applications, Nucl. Instrum. Meth. A, 504, 24–37, https://doi.org/10.1016/S0168-9002(03)00745-9, 2003. a, b

Liritzis, I., Stamoulis, K., Papachristodoulou, C., and Ioannides, K.: A re-evaluation of radiation dose-rate conversion factors, Mediterr. Archaeol. Ar., 13, 1–15, 2013. a

Løvborg, L. and Kirkegaard, P.: Response of 3′′× 3′′ NaI(Tl) detectors to terrestrial gamma radiation, Nucl. Instrum. Methods, 121, 239–251, https://doi.org/10.1016/0029-554x(74)90072-x, 1974. a, b, c, d

Martin, L., Incerti, S., and Mercier, N.: DosiVox: Implementing Geant 4-based software for dosimetry simulations relevant to luminescence and ESR dating techniques, Ancient TL, 33, 1–10, https://doi.org/10.26034/la.atl.2015.484, 2015. a

Martin, L., Duval, M., and Arnold, L. J.: To what extent do field conditions affect gamma dose rate determination using portable gamma spectrometry?, Radiat. Phys. Chem., 216, 111365, https://doi.org/10.1016/j.radphyschem.2023.111365, 2024. a

Mauz, B., Hubmer, A., Bahl, C., Lettner, H., and Lang, A.: Comparing two efficiency calibration methods used in gamma spectrometry, Ancient TL, 39, 12–17, https://doi.org/10.26034/la.atl.2021.552, 2021. a

Mercier, N. and Falguères, C.: Field gamma dose-rate measurement with a NaI(Tl) detector: re-evaluation of the technique, Ancient TL, 25, 1–4, https://doi.org/10.26034/la.atl.2007.400, 2007. a, b, c, d

Miallier, D., Guérin, G., Mercier, N., Pilleyre, T., and Sanzelle, S.: The Clermont radiometric reference rocks: a convenient tool for dosimetric purposes, Ancient TL, 27, 37–44, https://doi.org/10.26034/la.atl.2009.428, 2009. a, b, c, d, e, f, g

Murray, A., Bownman, S., and Aitken, M.: Evaluation of the gamma dose-rate contribution, PACT, 2, 84–96, 1978. a, b

Preusser, F. and Kasper, H. U.: Comparison of dose rate determination using high-resolution gamma spectrometry and inductively coupled plasma–mass spectrometry, Ancient TL, 19, 19–23, https://doi.org/10.26034/la.atl.2001.329, 2001. a, b

R Core Team: R: A language and environment for statistical computing, The R Foundation, https://r-project.org/ (last access: 19 May 2025), 2024. a

Rahman, R., Plater, A. J., Nolan, P. J., and Appleby, P. G.: Assessing CZT detector performance for environmental radioactivity investigations, Radiat. Prot. Dosim., 154, 477–482, https://doi.org/10.1093/rpd/ncs253, 2013. a

Rhodes, E. J. and Schwenninger, J. L.: Dose rates and radioisotope concentrations in the concrete calibration blocks at Oxford, Ancient TL, 25, 5–8, https://doi.org/10.26034/la.atl.2007.401, 2007. a, b

Richter, D., Dombrowski, H., Neumaier, S., Guibert, P., and Zink, A. C.: Environmental gamma dosimetry with OSL of -Al2O3:C for in situ sediment measurements, Radiat. Prot. Dosim., 141, 27–35, https://doi.org/10.1093/rpd/ncq146, 2010. a, b, c

Rieser, U.: Low-level gamma-spektrometrie zum zwecke der dosisleistungsbestimmung bei der lumineszenz datierung, Diploma thesis (unpublished), Ruprecht-Karls University of Heidelberg, Heidelberg, 1991. a, b, c, d

Sanderson, D. C. W. and Murphy, S.: Using simple portable OSL measurements and laboratory characterisation to help understand complex and heterogeneous sediment sequences for luminescence dating, Quat. Geochronol., 5, 299–305, https://doi.org/10.1016/j.quageo.2009.02.001, 2010. a

Scheiber, C. and Chambron, J.: CdTe detectors in medicine: a review of current applications and future perspectives, Nucl. Instrum. Meth. A, 322, 604–614, https://doi.org/10.1016/0168-9002(92)91239-6, 1992. a

Tudyka, K., Miłosz, S., Adamiec, G., Bluszcz, A., Poręba, G., Paszkowski, Ł., and Kolarczyk, A.: μDose_ A compact system for environmental radioactivity and dose rate measurement, Radiat. Meas., 118, 8–13, https://doi.org/10.1016/j.radmeas.2018.07.016, 2018. a

Tudyka, K., Kłosok, K., Gosek, M., Kolarczyk, A., Miłosz, S., Szymak, A., Pilśniak, A., Moska, P., and Poręba, G.: μDOSE+: Environmental radioactivity and dose rate measurement system with active shielding boosted by machine learning, Measurement, 234, 114854, https://doi.org/10.1016/j.measurement.2024.114854, 2024. a

Verger, L., Bonnefoy, J. P., Glasser, F., and Ouvrier-Buffet, P.: New Developments in CdTe and CdZnTe Detectors for X and y-Ray Applications, J. Electron. Mater., 26, 738–744, https://doi.org/10.1007/s11664-997-0225-2, 1997. a

Vermeesch, P.: IsoplotR: a free and open toolbox for geochronology, Geosci. Front., 9, 1479–1493, https://doi.org/10.1016/j.gsf.2018.04.001, 2018.  a

Wickham, H.: ggplot2: Elegant Graphics for Data Analysis, Springer-Verlag New York, https://doi.org/10.1007/978-3-319-24277-4, 2016. a

York, D., Evensen, N. M., Martínez, M. L., and De Basabe Delgado, J.: Unified equations for the slope, intercept, and standard errors of the best straight line, Am. J. Phys., 72, 367–375, https://doi.org/10.1119/1.1632486, 2004. a, b, c

Zöller, L. and Pernicka, E.: A note on overcounting in alpha-counters and its elimination, Ancient TL, 7, 11–14, https://doi.org/10.26034/la.atl.1989.140, 1989. a

1

The plus indicates a slightly higher energy resolution compared to the “non-plus” GR1 version.

Download
Short summary
Accurate readings on the environmental gamma dose rate are important. Portable gamma-ray detectors, such as those that are NaI- or LaBr3-based, are easy to handle and affordable. Limited information on alternatives, like CZT (cadmium zinc telluride) detectors, is available. We tested CZT detectors and found them suitable for in-field deployment. We used simulations and field tests to evaluate the optimal energy threshold for direct dose rate readings, making the CZT system a reliable alternative.
Share